CS-ECTA-0X is a test system specifically designed for medical fluorescence imaging systems. It enables quantitative evaluation of fluorescence concentration and detection depth. Featuring a flexible structure and adjustable fixture length, the system accommodates a wide range of endoscope types. With two dedicated testing modules for concentration and depth, it ensures highly repeatable testing under standardized conditions.
Application Background
Fluorescence imaging is increasingly applied in endoscopic surgeries, raising the demand for objective evaluation of system performance, such as sensitivity, signal-to-noise ratio (SNR), and stability.
The release of TGDMDMA 0031-2024 provides a unified benchmark for fluorescence endoscope performance testing.
CS-ECTA-0X was developed specifically to align with this standard and address typical industry requirements. It is ideal for use in endoscope manufacturing, medical device inspection, R&D validation, and quality control scenarios.
Core Product Features:
· Adjustable Fixture System: Supports secure fixation of endoscopes of various lengths and specifications to ensure stable imaging.
· Fluorescence Concentration Test Disk: Integrated with multiple fluorescence concentration standard samples for quick switching and comparison.
· Fluorescence Depth Test Slot: Designed with a standardized depth test structure to evaluate system detection capability.
· Standardized Test Workflow: Fully aligned with the TGDMDMA 0031-2024 standard requirements.
Supported Test Parameters (TGDMDMA 0031-2024 Standard)
Signal-to-Noise Ratio (SNR) |
A statistical analysis of variations in the output values of each pixel. |
Signal-to-Background Ratio (SBR) |
The ratio between the fluorescence signal and the background signal, reflecting the system’s ability to distinguish target from background. |
Sensitivity |
The minimum fluorescent agent concentration required for the imaging system to achieve effective visualization. |
Dynamic Range |
The ratio between the concentration at which the system reaches peak response and the minimum concentration defined by sensitivity. |
Signal Stability Interval |
The range of distances within which the fluorescence signal remains relatively stable in output intensity. |
Focal Plane Concordance Coefficient |
The ratio of the system’s response to high-contrast edges at the white-light focal plane versus the fluorescence focal plane. |
Relevant Standards
Standard Number | Standard Title | View Details |
TGDMDMA 0031-2024 |
Performance test methods for medical fluorescence imaging systems | View Details |
YY0068.1-2008 |
Medical endoscopes — Rigid Endoscope — Part 1: Optical properties and test methods | View Details |
YY/T 1603-2018 |
Medical endoscopes — Endoscope supply units-Video camera system | View Details |
YY1587-2018 |
Medical endoscopes — Video endoscopes | View Details |
【Reserve a Lab】For more information, please contact: sales@colorspace.com.cn
Follow ColorSpace for industry news/events/solutions
Contact: sales@colorspace.com.cn
Phone: 15817270587
Tel: 400-886-3881(周一至周五 9:00~18:00)
Email: sales@colorspace.com.cn
Add: South Building #2-1101, Lv Di Qi Hang Office Building, Hou Sha Yu, Shun Yi District, Beijing, China