The CS-ECTA-1X is a professional fluorescence performance testing system designed specifically for medical endoscopes. It evaluates critical fluorescence metrics including response characteristics, signal-to-noise ratio (SNR), and image latitude with high stability and repeatability. The system integrates a dedicated fluorescence testing platform and an automated inspection unit to deliver accurate, quantitative results. Fully compliant with the latest medical imaging system draft standards, it is ideal for endoscope R&D validation, quality control, and image optimization processes.
Why Choose CS-ECTA-1X?
· Comprehensive Testing: Measures key fluorescence parameters to ensure superior image quality.
· Automated & Reliable: Automated detection for consistent and repeatable testing results.
· Standard Compliant: Meets or exceeds requirements in current medical endoscope fluorescence testing standards.
· Versatile Applications: Supports product development, quality assurance, and regulatory submissions.
Application Background
Fluorescence imaging plays an increasingly vital role in endoscopic surgeries, requiring stricter image quality control to ensure diagnostic accuracy. Key performance indicators such as response characteristics, SNR, and image latitude directly impact image clarity and reliability. The CS-ECTA-1X system is developed to meet these critical testing needs, helping manufacturers efficiently achieve standardized and reliable fluorescence performance evaluation.
· Dedicated Testing Fixture Platform: Robust and stable structure designed to accommodate a wide range of endoscope models and configurations.
· Automated Fluorescent Solution Testing Carousel: Equipped with multiple fluorescent concentration standards for seamless automatic switching and testing.
· Specialized High-Uniformity Integrating Sphere System: Provides highly consistent illumination with minimal stray light background, ensuring precise and accurate measurements.
Supported Testing Parameters
Parameter |
Description |
Fluorescence Response Characteristics |
Outputs electro-optical conversion functions or data lists with linear fit coefficient R² ≥ 0.98. For nonlinear systems, corresponding evaluation methods must be provided. |
Fluorescence Signal-to-Noise Ratio (SNR) |
Provides nominal values and imaging modes, with a tolerance range of -20% (no upper limit). |
Fluorescence Static Image Latitude |
Provides nominal values and imaging modes, with a tolerance range of -20% (no upper limit). |
【Reserve a Lab】For more information, please contact: sales@colorspace.com.cn
Follow ColorSpace for industry news/events/solutions
Contact: sales@colorspace.com.cn
Phone: 15817270587
Tel: 400-886-3881(周一至周五 9:00~18:00)
Email: sales@colorspace.com.cn
Add: South Building #2-1101, Lv Di Qi Hang Office Building, Hou Sha Yu, Shun Yi District, Beijing, China