The medical industry uses imaging technology in many ways, such as endoscopes, pill cameras, head-mounted cameras, and so on. Therefore, the quality testing of image products is also essential. Medical cameras are often used to capture smaller objects, so small size test cards are used, usually with transmissive light sources. In some cases, uniform reflection light sources or light boxes are also used to analyze color. Angular resolution, viewing angle, effective depth of field, color resolution, color reproduction and distortion are the test parameters that are often concerned.
Common test items:
·Total number of optically active pixels | ·Angular resolution | ·Color resolution and color reproducibility | ·Lens test parameters |
·Angle of view | ·Effective depth range | ·Light energy transfer efficiency-effective luminosity | ·Signal-to-noise ratio |
·Angle of view | ·Rate of illumination change | ·Lighting effectiveness | ·Comprehensive light efficiency |
·Static image tolerance |
·Unit relative distortion | ·Focus | ·Brightness response characteristics |
YY 1298-2016 Medical endoscopes Capsule endoscopes
YY/T 1587-2018 Medical endoscopes
YY/T 1603-2018: Medical endoscopes Endoscope function supply units Camera systems
YY 0068.1-2008 Medical endoscopes-Rigid endoscopes-Part 1: Optical properties and test methods
YY 1081-2011: Medical endoscopes Endoscope function supply units Cold light sources
GB/T 36480-2018 General specification for compact embedded cameras in information technology
GB 11244-2005-General requirements for medical endoscopes and accessories
More Industry Standards → [Medical Camera Testing Industry Standards]
Test equipment |
Model |
Configuration/Test Items |
software |
||
Medical Camera Imaging Quality Analysis System V3.0 |
- |
Image quality analysis software supporting multiple medical industry standards |
imatest master |
Image quality analysis software |
|
standard light source system |
||
standard fiber optic system |
ECTS-SP02 |
Standard lighting conditions are provided for testing the spectral distribution and color reproducibility of optical mirrors |
integrating sphere |
ECTS-SP01 |
For analyzing spectral power distribution of optical mirrors, color reproduction |
Light Booth |
LE006-D-L |
For testing luminance response function, provide illumination environment reflecting standard gray card |
testing equipment |
||
CS-ECTS |
Used for medical endoscope hard scope, soft scope, electronic scope, capsule endoscope and other systems imaging performance test |
|
CS-ECTS-04 |
Performance of standard test light sources for testing medical cold light sources (YYT1081) |
|
luminance meter |
CS-160 |
Used to test luminance response function |
Spectral photometer |
CL-500A |
Measure the color temperature, illuminance, common range, spectral curve and color rendering index of the light source |
T-10A |
Small probe for testing the edge uniformity of the lens body |
|
Spectrophotometer |
CS-i1pro |
Display color reproduction for testing electronic mirrors |
MTF and Focus Test Equipment |
CS-ECTS-03 |
Testing the focal length of a detachable optic |
Test Chart |
||
ECTS-TC01 |
luminance response function, signal-to-noise ratio, tolerance |
|
ECTS-TC02 |
definition test |
|
ECTS-TC03 |
Clarity test, depth of field |
|
ECTS-TC04 |
Clarity test, depth of field |
|
ECTS-TC05 |
angular resolution test |
|
ECTS-TC06 |
Clarity test, distortion |
|
ECTS-TC14 |
Signal-to-noise ratio, dynamic range test |
|
ECTS-TC07 |
Distortion, transverse chromatic aberration |
|
ECTS-TC13 |
Field position correction, uniformity test |
|
ECTS-TC08 |
Color reproduction, color difference test |
|
ECTS-TC09 |
Color reproduction, color difference test |
|
ECTS-TC15 |
definition test |
|
distortion test chart |
ECTS-TC10 |
Distortion test card, conforming to YY0068.1 standard, used for testing distortion |